Sensor Alignment and Calibration
IM-Mercury_II_6000 Series Rev. 1 Page 14 ©2014 MicroE Systems
5.2 Using the Software
Use the following instructions to perform sensor alignment and calibration using the Alignment
Tool and the SmartPrecision II Software.
5.2.1 Connect the Alignment Tool and Encoder
Step
Action
1.
Verify that the Alignment Tool Adapter/SmartPrecision II Alignment Tool is not powered on. The
Power/Cal. LED should be off.
2.
Connect the MII 6000 encoder and Ethernet cable as shown below.
3.
Power up the Alignment Tool by plugging in 5VDC to the Alignment Tool Adapter.
Result: The Power/Cal. LED illuminates a steady green. The sensor is initialized after an
approximate 10 second delay.
5VDC Power
Connector
Alignment Tool Adapter
Power/Cal. LED
Power/Cal. LED
D
Alignment Tool
Encoder
Ethernet Cable
Sensor Connector
Computer
(Windows or
other OS)